Citation
Alsayed, Abdallah and Raja Ahmad, Raja Mohd Kamil and Ramli, Hafiz Rashidi and As'arry, Azizan
(2019)
Design and calibration of pinch force measurement using strain gauge for post-stroke patients.
International Journal of Integrated Engineering, 11 (4).
pp. 40-50.
ISSN 2229-838X; ESSN: 2600-7916
Abstract
Two fingers strength is an indicative measurement of pinch impairment. Conventionally, Fugl Meyer Upper Extremity Assessment (FMA-UE) is the primary standard to measure pinch strength of post-stroke survivors. In literature, the evaluation method performed by the therapist is subjective and exposed to inter-rater and intra-rater reliabilities. Recently, force-sensing resistors were implemented to measure two fingers force, but these sensors are subjected to nonlinearity, high hysteresis, and voltage drift. This paper presents a design of pinch force measurement based on the strain gauge. The pinch sensor was calibrated within a range of between 0 N to 50 N over a pinching length of 20 mm with a linearity error of 0.0123% and hysteresis of 0.513% during the loading and unloading process. The voltage drift has an average of 0.24% over 20 minutes. The pinch force measurement system reveals an objective pinch force measurements in evaluating the rehabilitation progress of post-stroke patients.
Download File
Additional Metadata
Actions (login required)
|
View Item |