UPM Institutional Repository

Low power upconversion mixer for medical remote sensing


Citation

Lioe, De Xing and Shafie, Suhaidi and Ramiah, Harikrishnan and Tan, Gim Heng (2014) Low power upconversion mixer for medical remote sensing. The Scientific World Journal, 2014. art. no. 923893. pp. 1-5. ISSN 2356-6140; ESSN: 1537-744X

Abstract

This work presents the design of a low power upconversion mixer adapted in medical remote sensing such as wireless endoscopy application. The proposed upconversion mixer operates in ISM band of 433 MHz. With the carrier power of −5 dBm, the proposed mixer has an output inferred 1 dB compression point of −0.5 dBm with a corresponding output third-order intercept point (OIP3) of 7.1 dBm. The design of the upconversion mixer is realized on CMOS 0.13 μm platform, with a current consumption of 594 μA at supply voltage headroom of 1.2 V.