Analysis of Low Lead and Cadmium Concentrations in Soil Extracts by Graphite Furnace Atomic Absorption Spectrophotometry


Citation

Jopony, Marcus (1993) Analysis of Low Lead and Cadmium Concentrations in Soil Extracts by Graphite Furnace Atomic Absorption Spectrophotometry. Pertanika Journal of Science & Technology, 1 (2). pp. 199-207. ISSN 0128-7680

Abstract

An analytical procedure was derived to provide an accurate measurement of low lead (Pb) and cadmium (Cd) concentrations in soil extracts using graphite furnace atomic absorption spectrometry (GFAAS). Based on peak area measurements and with background correction, the use of 2% phosphoric acid as chemical modifier enables the GFAAS to give measurements of Pb and Cd concentration as low as 3 fLg dm-3 with a high degree of precision. The modifier ensures a single well-defined atomization peak of the analyte, and therefore higher sensitivity and accuracy in the measurements.


Download File

[img]
Preview
PDF
Analysis_of_Low_Lead_and_Cadmium_Concentrations_in.pdf

Download (2MB)

Additional Metadata

Item Type: Article
Publisher: Universiti Putra Malaysia Press
Keywords: cadmium, chemical modifier, GFAAS, lead, soil extracts
Depositing User: Nur Izzati Mohd Zaki
Date Deposited: 23 Nov 2009 06:41
Last Modified: 27 May 2013 07:05
URI: http://psasir.upm.edu.my/id/eprint/3068
Statistic Details: View Download Statistic

Actions (login required)