Performance characterisation has been carried out of nanoDot™ OSLDs and TLD-100™, validating their use for diagnostic x-ray dose profile assessments. Investigations include optical annealing, signal depletion, signal fading, dose-response, sensitivity, and energy dependence. Both dosimeter types were exposed using the Constant Potential Industrial X-ray (Model Philips MG 165) in Standard Radiation Qualities (RQT) procedure located in Nuclear Malaysia, Bangi. OSLD annealing using a 14 W compact fluorescent lamp showed an average signal loss of ∼93% as a result of 60 min of illumination. For dosimetric signal depletion, screened nanoDot™ OSLDs gave rise to the most favorable performance, with a mean signal loss of 1.0% per reading. In respect of signal fading, similar favorable performance was found for the screened nanoDot™ OSLDs, after a stabilisation period of 11–12 days post-irradiation, the average reading decreased by ∼1% over a further 17 days. For doses up to 500 mGy the TLD-100™ and screened nanoDot™ OSLDs both provide a highly linear response, with a regression coefficient of 0.999 in both cases. Linear energy dependence was found for RQT spectra from 100 to 150 kV.
Download File
Full text not available from this repository.
Official URL or Download Paper: https://www.sciencedirect.com/science/article/pii/...
|
Additional Metadata
Item Type: | Article |
---|---|
Divisions: | Faculty of Science |
DOI Number: | https://doi.org/10.1016/j.radphyschem.2022.110365 |
Publisher: | Elsevier |
Keywords: | Characterisation; CT; OSLD; RQT |
Depositing User: | Ms. Nur Faseha Mohd Kadim |
Date Deposited: | 18 Jul 2024 07:29 |
Last Modified: | 18 Jul 2024 07:29 |
URI: | http://psasir.upm.edu.my/id/eprint/100161 |
Statistic Details: | View Download Statistic |
Actions (login required)
View Item |