A critical appraisal of system-reported Organ Dose (OD) versus manually calculated Mean Glandular Dose (MGD) in Dubai’s mammography services


Citation

Mohammad Noor, Kaltham Abdulwahid and Mohd Norsuddin, Norhashimah and Abdul Karim, Muhammad Khalis and Che Isa, Iza Nurzawani and Ulaganathan, Vaidehi (2025) A critical appraisal of system-reported Organ Dose (OD) versus manually calculated Mean Glandular Dose (MGD) in Dubai’s mammography services. Diagnostics, 15 (1). art. no. 81. pp. 1-13. ISSN 2075-4418

Abstract

Background: This study compares system-reported organ doses (ODs) to manually calculated mean glandular doses (MGDs) in mammography across multiple centers and manufacturers in Dubai. Methods: A retrospective study of 2754 anonymized mammograms from six clinics in Dubai were randomly retrieved from a central dose survey database. Organ doses were documented along with other dosimetry information like kVp, mAs, filter, target, compression force, and breast thickness. Mean glandular doses, MGDs, were calculated manually for all the patients using the Dance formula and inferential statistical analyses were run to compare the two figures and verify the factors affecting each. Results: Our study’s analysis revealed that manually calculated mean glandular doses (MGDs) provide a more reliable indicator of radiation exposure than organ doses (ODs) reported by DICOM, particularly in multi-vendor scenarios. Manually calculated MGD values were consistently lower than system-reported ODs (MLO view: 0.96 ± 0.37 mGy vs. 1.38 ± 0.45 mGy; CC view: 0.81 ± 0.33 mGy vs. 1.22 ± 0.38 mGy). Significant differences in both system-reported ODs and manually calculated MGDs were observed across centers (p < 0.001). Strong correlations between system-reported ODs and manually calculated MGDs were found for Siemens equipment (r = 0.923, p < 0.001) but only moderate correlations for GE systems (r = 0.638, p < 0.001). Calculated MGD values were significantly higher for GE equipment compared to Siemens (1.49 ± 0.77 mGy vs. 0.93 ± 0.33 mGy, p < 0.001). Conclusions: This study addresses concerns regarding mammography dosimetry accuracy by demonstrating the superiority of mean glandular doses over DICOM-generated organ doses. These findings empower practitioners to optimize dose levels, ensuring safer and more effective breast cancer screening protocols.


Download File

[img] Text
123469.pdf - Published Version
Available under License Creative Commons Attribution.

Download (3MB)
Official URL or Download Paper: https://www.mdpi.com/2075-4418/15/1/81

Additional Metadata

Item Type: Article
Subject: Clinical Biochemistry
Divisions: Faculty of Science
DOI Number: https://doi.org/10.3390/diagnostics15010081
Publisher: Multidisciplinary Digital Publishing Institute (MDPI)
Keywords: Diagnostic reference level; Mammography; Mean glandular dose; Organ dose; Radiation dose
Sustainable Development Goals (SDGs): SDG 3: Good Health and Well-being, SDG 9: Industry, Innovation and Infrastructure, SDG 12: Responsible Consumption and Production
Depositing User: Ms. Siti Radziah Mohamed@mahmod
Date Deposited: 22 Apr 2026 12:12
Last Modified: 22 Apr 2026 12:12
Altmetrics: http://www.altmetric.com/details.php?domain=psasir.upm.edu.my&doi=10.3390/diagnostics15010081
URI: http://psasir.upm.edu.my/id/eprint/123469
Statistic Details: View Download Statistic

Actions (login required)

View Item View Item