Crouse, Karen Anne (1983) A calorimetric technique of measuring low heat of solution. Pertanika, 6 (1). pp. 93-94.
| PDF 775Kb |
Abstract
A simple calorimeter has been assembled specially for the measurement of low heats of solution. Decanoic acid was used as a thermostatic material. A change in temperature induces an emf in a copper constantan thermocouple. This small voltage is measured with a microvoltmeter, amplified and recorded
| Item Type: | Article |
|---|---|
| ID Code: | 2139 |
| Deposited By: | Nur Izzati Mohd Zaki |
| Deposited On: | 11 Nov 2009 09:08 |
| Last Modified: | 22 Jul 2010 22:51 |